Physio-Control LIFEPAK 1000 AED
In a squad car, onboard a fire engine or out in the field, you're always ready to respond to sudden cardiac arrest emergencies. The LIFEPAK 1000 provides powerful defibrillation, long battery life and robust durability that can stand up to severe environments.
The LIFEPAK 1000 combines the simplicity of an AED with clear guidance, including onscreen and audio prompts with the flexibility of manual mode operation when indicated. Flexible setup options allow on-screen graphics for the basic rescuer or ECG display and access to manual mode for the more advanced rescuer. This powerful AED combines basic guidance with advanced operational control, enabling responders to deliver fast, appropriate treatment.
- Exclusive cprMAX technology helps minimize CPR interruptions by allowing responders to perform pre-shock CPR while the AED charges.
- Industry-leading features including a CPR countdown timer, ECG capability, a shock counter and programmable settings – all on a large, easy-to-read LCD screen.
- Loud voice prompts and on-screen graphics provide clear guidance on applying electrodes and initiating a shock for added user confidence.
- Escalating energy to 360J biphasic for difficult to defibrillate patients.
- Provides battery power for up to 220 shocks – escalating to 360 joules – delivering 17 hours of monitoring time.
- Compatibility with other LIFEPAK electrodes and monitors speeds the transition to the next level of care.
- Post-event patient and performance data can be easily viewed, analyzed and managed with CODE-STAT data review software – helping to improve future response and target training needs.
- Endures rigorous drop-tests and includes protective case and bumpers to help maximize durability in the harshest environments.
- Designed to ride along in any vehicle without damage from continuous vibrations and other movement.
Waveform |
Biphasic truncated exponential with voltage and duration compensation for patient impedance |
Energy sequences |
2User configurable, 150 joules-360 joules. Default energy output settings are 200, 300, 360 joules. 360 joules for every shock thereafter |
Flexible energy |
Increases only after a lower energy was unsuccessful. |
Charge time |
With new, nonrechargeable battery pack; 200 joules in less than 9 seconds (360 joules in less than 15 seconds) |
3-Wire (Lead II) monitoring capability |
If ECG display option purchased). Requires purchase of 3-wire (Lead II) monitoring cable and LIFE-PATCH® electrodes |
Device software |
Field upgradeable |
Infant/Child reduced energy defibrillation electrodes |
Reduces selected energy by a factor of 4. Intended for use only with children up to 8 years of age or 55 lbs (25 kg) |
Safety classification |
Internally powered equipment IEC 60601-1 |
Electrical protection |
Input protected against high voltage defibrillator pulses per IEC 60601-1 |
Modes |
- AED: Provides operating capability for basic users. |
Controls |
On/Off, Shock, Menu, Two (2) configurable soft keys |
Auto analyze |
User can configure device to auto analyze, auto analyze after first shock, or prompt user to push analyze key before each analysis period. |
CPR time |
(Post shock or after no shock advised) User configurable - 15, 30, 45, 60, 90, 120, 180 seconds |
Voice prompt volume |
Allows user to change speaker volume |
ECG display |
(If option purchased) - Turns display on/off for AED mode. |
Motion detection |
User defined On/Off (default On). |
Service alert |
Audio alarm if the device needs servicing. Configurable on/off. |
Manual access |
(If ECG display option purchased) - Devices configured with an ECG display may be set up to allow user to initiate a charge and shock without analysis. |
Initial CPR |
User defined time for CPR after first analysis regardless of analysis decision. Can be set to OFF, 15, 30, 45, 60, 90, 120 and 180 seconds. |
Pre-shock CPR |
Allows for CPR while device is charging. Can be set to OFF, 15, or 30 seconds. |
Stacked shocks |
(ON/OFF) When Off, allows for provision of CPR after each shock. |
Pulse check |
(Always, After Every NSA, After Second NSA, Never) Allows device to prompt for a pulse check either after each shock, after every NSA pulse check, or never prompt for a pulse check (default Never). |
Display |
Backlit LCD displays number of shocks delivered, elapsed time, text and graphics of heart rhythm and optional ECG |
Nonrechargeable battery |
- Type: Lithium manganese dioxide (Li/MnO2), 12.0 V, 4.5 Ah |
Rechargeable battery |
Battery |